Lab statistics
Q1Lab statistics · Easy
The sum of all control values divided by the number of values is the:
AA: Mean
BB: Mode
CC: Range
DD: Median
Show answer
Answer: A — Mean
The median is the middle value; the mode is the most frequent value; the range is highest minus lowest.
Reference: Tietz Fundamentals of Clinical Chemistry
Q2Lab statistics · Easy
Which statistic measures how widely values are spread around the mean?
AA: Mean
BB: Standard deviation
CC: Median
DD: Sensitivity
Show answer
Answer: B — Standard deviation
SD is the square root of the variance. A small SD means good precision.
Reference: Tietz Fundamentals of Clinical Chemistry
Q3Lab statistics · Easy
The variance of a control is 144. The standard deviation is:
AA: 12
BB: 14.4
CC: 72
DD: 144
Show answer
Answer: A — 12
SD = square root of variance = square root of 144 = 12.
Reference: Tietz Fundamentals of Clinical Chemistry
Q4Lab statistics · Easy
The coefficient of variation (CV%) is calculated as:
AA: SD x mean
BB: Variance / mean
CC: Mean / SD x 100
DD: SD / mean x 100
Show answer
Answer: D — SD / mean x 100
CV expresses precision as a percentage, so methods at different concentrations can be compared.
Reference: Tietz Fundamentals of Clinical Chemistry
Q5Lab statistics · Medium
A control has a mean of 200 mg/dL and an SD of 6 mg/dL. The CV is:
AA: 1.2%
BB: 3%
CC: 6%
DD: 33%
Show answer
Answer: B — 3%
6 / 200 x 100 = 3%.
Reference: Tietz Fundamentals of Clinical Chemistry
Q6Lab statistics · Medium
A control has a mean of 38.5 and an SD of 4.2. Its 2SD acceptable range is:
AA: 34.3 - 42.7
BB: 30.1 - 46.9
CC: 25.9 - 51.1
DD: 36.4 - 40.6
Show answer
Answer: B — 30.1 - 46.9
Mean +/- 2SD = 38.5 +/- 8.4. Mean +/- 1SD (34.3 - 42.7) is too narrow to use as the acceptable range.
Reference: Westgard: Basic QC Practices
Q7Lab statistics · Medium
CLSI recommends that a laboratory establish a new reference interval by testing at least:
AA: 20 healthy people
BB: 120 healthy people per group
CC: 1,000 patients
DD: 10 hospital patients
Show answer
Answer: B — 120 healthy people per group
A lab can verify an existing (transferred) interval with about 20 healthy people. The reference interval usually covers the central 95% of results.
Getting the same result on repeated analysis of one sample describes:
AA: Sensitivity
BB: Specificity
CC: Accuracy
DD: Precision (reproducibility)
Show answer
Answer: D — Precision (reproducibility)
Accuracy is closeness to the true value. A method can be precise but inaccurate.
Reference: Tietz Fundamentals of Clinical Chemistry
Q9Lab statistics · Medium
The ability of a test to correctly identify people who have the disease is:
AA: Sensitivity
BB: Positive predictive value
CC: Precision
DD: Specificity
Show answer
Answer: A — Sensitivity
Sensitivity = TP / (TP + FN). Specificity = TN / (TN + FP) and correctly identifies people without the disease.
Reference: Tietz Fundamentals of Clinical Chemistry
Q10Lab statistics · Hard
The positive predictive value of a test increases when:
AA: Disease prevalence decreases
BB: Disease prevalence increases
CC: Sensitivity falls
DD: The test is repeated on the same sample
Show answer
Answer: B — Disease prevalence increases
PPV depends on prevalence; screening low-prevalence populations gives more false positives.
Reference: Tietz Fundamentals of Clinical Chemistry
QC principles
Q11QC principles · Easy
Material with a precisely known concentration used to calibrate an analyzer is a:
AA: Patient pool
BB: Control
CC: Standard (calibrator)
DD: Blank
Show answer
Answer: C — Standard (calibrator)
Controls monitor performance; calibrators (standards) set the relationship between signal and concentration.
Reference: Tietz Fundamentals of Clinical Chemistry
Q12QC principles · Easy
A Levey-Jennings chart plots:
AA: Absorbance against wavelength
BB: Sensitivity against specificity
CC: Patient results against age
DD: Control results over time against the mean and SD limits
Show answer
Answer: D — Control results over time against the mean and SD limits
Lines are drawn at the mean and +/-1, 2 and 3 SD.
Reference: Westgard: Basic QC Practices
Q13QC principles · Medium
Six or more consecutive control values on one side of the mean, after an abrupt change, is called a:
AA: Shift
BB: Random error
CC: Outlier
DD: Trend
Show answer
Answer: A — Shift
Common causes: new reagent or calibrator lot, recalibration, or a change in instrument settings.
Reference: Westgard: Basic QC Practices
Q14QC principles · Medium
Control values moving steadily in one direction over several runs is a:
AA: Shift
BB: Trend
CC: Random error
DD: Delta failure
Show answer
Answer: B — Trend
Common causes: gradual reagent or calibrator deterioration, aging lamp, slow temperature change or dirty optics. Find and fix the cause; don't just change reagents by habit.
Reference: Westgard: Basic QC Practices
Q15QC principles · Medium
Errors that are predictable and push all results in one direction (bias) are:
AA: Pre-analytical errors only
BB: Random errors
CC: Systematic errors
DD: Clerical errors
Show answer
Answer: C — Systematic errors
Systematic error affects accuracy and is usually fixed by recalibration or reagent correction. Random error affects precision.
A result is above the analyzer's measuring range. The correct action is:
AA: Dilute the sample, re-run, and multiply by the dilution factor
BB: Report as 'high' with no value always
CC: Recalibrate
DD: Report as is
Show answer
Answer: A — Dilute the sample, re-run, and multiply by the dilution factor
Check the dilution is validated for that analyte.
Reference: Tietz Fundamentals of Clinical Chemistry
Calculations
Q30Calculations · Easy
In a twofold serial dilution starting at 1:2, the sixth tube is:
AA: 1:12
BB: 1:32
CC: 1:64
DD: 1:128
Show answer
Answer: C — 1:64
1:2, 1:4, 1:8, 1:16, 1:32, 1:64.
Reference: Tietz Fundamentals of Clinical Chemistry
Q31Calculations · Easy
To prepare 100 mL of a 1% (w/v) albumin solution you need:
AA: 0.1 g
BB: 1 g
CC: 10 g
DD: 100 g
Show answer
Answer: B — 1 g
1% w/v = 1 g per 100 mL.
Reference: Tietz Fundamentals of Clinical Chemistry
Q32Calculations · Easy
Moles of solute per liter of solution is:
AA: Normality
BB: Osmolality
CC: Molality
DD: Molarity
Show answer
Answer: D — Molarity
Molality is moles per kilogram of solvent.
Reference: Tietz Fundamentals of Clinical Chemistry
Q33Calculations · Hard
Total calcium is 2.10 mmol/L and albumin 31 g/L. Using corrected Ca = Ca + 0.02 x (40 - albumin), the corrected calcium is:
AA: 2.12 mmol/L
BB: 2.19 mmol/L
CC: 2.28 mmol/L
DD: 2.50 mmol/L
Show answer
Answer: C — 2.28 mmol/L
2.10 + 0.02 x 9 = 2.10 + 0.18 = 2.28 mmol/L. Ionized calcium is better when accuracy matters.
Reference: Tietz Fundamentals of Clinical Chemistry
Q34Calculations · Medium
To convert BUN (mg/dL) to urea (mg/dL), multiply by:
AA: 0.467
BB: 1.0
CC: 2.14
DD: 6.0
Show answer
Answer: C — 2.14
Urea (60 g/mol) contains 28 g of nitrogen: 60/28 = 2.14.
Reference: Tietz Fundamentals of Clinical Chemistry
Q35Calculations · Medium
Using a standard in a photometric assay, the sample concentration is:
AA: Absorbance x wavelength
BB: Sample absorbance x standard absorbance
CC: (Sample absorbance / standard absorbance) x standard concentration
DD: Standard concentration / sample absorbance
Show answer
Answer: C — (Sample absorbance / standard absorbance) x standard concentration
Based on Beer-Lambert law: absorbance is proportional to concentration.
Reference: Tietz Fundamentals of Clinical Chemistry
Lab water & errors
Q36Lab water & errors · Medium
Ion-exchange resins are used to produce:
AA: Tap water
BB: Saline
CC: Distilled water
DD: Deionized water
Show answer
Answer: D — Deionized water
Distillation uses boiling and condensation. Most analyzers need clinical laboratory reagent water (CLRW, formerly Type I), as specified by the manufacturer.
Reference: Tietz Fundamentals of Clinical Chemistry
Q37Lab water & errors · Easy
Most laboratory errors occur in which phase?
AA: Pre-analytical
BB: Post-analytical
CC: Equally in all phases
DD: Analytical
Show answer
Answer: A — Pre-analytical
Examples: wrong patient ID, wrong tube, hemolysis, delays. Post-analytical errors include transcription and reporting errors.
Reference: Tietz Fundamentals of Clinical Chemistry
Q38Lab water & errors · Medium
High total protein with low sodium measured by indirect ISE suggests:
AA: True hyponatremia
BB: Pseudohyponatremia
CC: Hemolysis
DD: EDTA contamination
Show answer
Answer: B — Pseudohyponatremia
Very high lipids or proteins reduce the water fraction of plasma. Direct ISE (e.g., blood gas analyzer) gives the true value.
Reference: Tietz Fundamentals of Clinical Chemistry